Characterization of Effective Orifice Areas of Mitral Prosthetic Heart Valves: An In-Vitro Study. 2017

Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
Aix-Marseille Université, CNRS, Marseille, France.

Reference values of hemodynamic parameters for the assessment of prosthetic heart valves are necessary, and ideally need to be provided by entities independent of the valve manufacturers. Thus, the study aim was to provide, in vitro, normal reference values of the effective orifice area (EOA) for different models and sizes of mitral prosthetic valve, and to assess the determinants of EOA and mean transvalvular pressure gradient (mTPG). Four models of mechanical prostheses were tested (one mono-leaflet, three bi-leaflet) and four models of bioprostheses (two bovine pericardial, two porcine) on a double-activation pulsed duplicator that was specifically designed and optimized for assessing the hemodynamic performance of mitral prosthetic valves. The hemodynamic conditions were standardized and included for bioprostheses: two mitral flow volumes, three mean aortic pressures, two heart rates, and three E/A ratios. The EOAs were measured with Doppler echocardiography, using the same method (continuity equation) as was used in the clinical setting. Overestimation in term of EOA was defined according to guidelines as >0.25 cm2. EOA reference values were recorded. For mono-leaflet prostheses (Medtronic Hall 7700, size 25 to 31 mm) 2.29 and 3.49; for bi-leaflet prostheses (St. Jude Medical Master and Master HP, sizes 25 to 33 mm and On-X valve, sizes 27-29 mm) 1.34 and 4.74 cm2; for porcine bioprostheses (Medtronic Mosaic CINCH, sizes 25 to 31 mm and St. Jude Epic 100, sizes 25 to 33 mm) 1.35 and 3.56 cm2; for bovine pericardial bioprosthetic valves (Edwards Perimount 6900P and Magna Ease 7300, sizes 25 to 33 mm) 1.67 and 2.36 cm2. There were some discrepancies between the normal reference EOAs measured compared to those provided by the prosthesis manufacturers, or in published reports. The bioprosthetic EOAs were shown to be smaller than the manufacturers' values in 32% of valves (by an average of 0.57 ± 0.28 cm2) versus in 7% of valves when compared to values reported elsewhere (by an average of 0.43 ± 0.17 cm2). The relationship between EOA and internal orifice area (IOA) varied according to the type of prosthesis. The EOA was close to the IOA in mechanical valves (regression slopes 0.87-0.99) but was much smaller than the IOA in bioprosthetic valves (slopes 0.25-0.30). The EOA was influenced by prosthesis diameter, prosthesis stent diameter and height, while the mTPG was influenced by EOA and heart rate. The present study has provided normal reference values of EOAs for several frequently used mitral prostheses. This information may be helpful for identifying and quantifying prosthetic valve dysfunction and prosthesis-patient mismatch.

UI MeSH Term Description Entries
D008422 Materials Testing The testing of materials and devices, especially those used for PROSTHESES AND IMPLANTS; SUTURES; TISSUE ADHESIVES; etc., for hardness, strength, durability, safety, efficacy, and biocompatibility. Biocompatibility Testing,Biocompatible Materials Testing,Hemocompatibility Testing,Testing, Biocompatible Materials,Testing, Hemocompatible Materials,Hemocompatibility Testings,Hemocompatible Materials Testing,Materials Testing, Biocompatible,Materials Testing, Hemocompatible,Testing, Biocompatibility,Testing, Hemocompatibility,Testing, Materials,Testings, Biocompatibility
D008943 Mitral Valve The valve between the left atrium and left ventricle of the heart. Bicuspid Valve,Bicuspid Valves,Mitral Valves,Valve, Bicuspid,Valve, Mitral,Valves, Bicuspid,Valves, Mitral
D011474 Prosthesis Design The plan and delineation of prostheses in general or a specific prosthesis. Design, Prosthesis,Designs, Prosthesis,Prosthesis Designs
D012016 Reference Values The range or frequency distribution of a measurement in a population (of organisms, organs or things) that has not been selected for the presence of disease or abnormality. Normal Range,Normal Values,Reference Ranges,Normal Ranges,Normal Value,Range, Normal,Range, Reference,Ranges, Normal,Ranges, Reference,Reference Range,Reference Value,Value, Normal,Value, Reference,Values, Normal,Values, Reference
D006350 Heart Valve Prosthesis A device that substitutes for a heart valve. It may be composed of biological material (BIOPROSTHESIS) and/or synthetic material. Prosthesis, Heart Valve,Cardiac Valve Prosthesis,Cardiac Valve Prostheses,Heart Valve Prostheses,Prostheses, Cardiac Valve,Prostheses, Heart Valve,Prosthesis, Cardiac Valve,Valve Prostheses, Cardiac,Valve Prostheses, Heart,Valve Prosthesis, Cardiac,Valve Prosthesis, Heart
D006439 Hemodynamics The movement and the forces involved in the movement of the blood through the CARDIOVASCULAR SYSTEM. Hemodynamic
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D001705 Bioprosthesis Prosthesis, usually heart valve, composed of biological material and whose durability depends upon the stability of the material after pretreatment, rather than regeneration by host cell ingrowth. Durability is achieved 1, mechanically by the interposition of a cloth, usually polytetrafluoroethylene, between the host and the graft, and 2, chemically by stabilization of the tissue by intermolecular linking, usually with glutaraldehyde, after removal of antigenic components, or the use of reconstituted and restructured biopolymers. Glutaraldehyde-Stabilized Grafts,Heterograft Bioprosthesis,Porcine Xenograft Bioprosthesis,Xenograft Bioprosthesis,Bioprostheses,Bioprostheses, Heterograft,Bioprostheses, Porcine Xenograft,Bioprostheses, Xenograft,Bioprosthesis, Heterograft,Bioprosthesis, Porcine Xenograft,Bioprosthesis, Xenograft,Glutaraldehyde Stabilized Grafts,Glutaraldehyde-Stabilized Graft,Graft, Glutaraldehyde-Stabilized,Grafts, Glutaraldehyde-Stabilized,Heterograft Bioprostheses,Porcine Xenograft Bioprostheses,Xenograft Bioprostheses,Xenograft Bioprostheses, Porcine,Xenograft Bioprosthesis, Porcine
D015150 Echocardiography, Doppler Measurement of intracardiac blood flow using an M-mode and/or two-dimensional (2-D) echocardiogram while simultaneously recording the spectrum of the audible Doppler signal (e.g., velocity, direction, amplitude, intensity, timing) reflected from the moving column of red blood cells. Doppler Echocardiography,Echocardiography, Continuous Doppler,Echocardiography, Two-Dimensional Doppler,2-D Doppler Echocardiography,2D Doppler Echocardiography,Continuous Doppler Echocardiography,Doppler Echocardiography, 2-D,Doppler Echocardiography, 2D,Doppler Echocardiography, Continuous,Doppler Echocardiography, Two-Dimensional,Echocardiography, 2-D Doppler,Echocardiography, 2D Doppler,Two-Dimensional Doppler Echocardiography,2 D Doppler Echocardiography,Doppler Echocardiography, 2 D,Doppler Echocardiography, Two Dimensional,Echocardiography, 2 D Doppler,Echocardiography, Two Dimensional Doppler,Two Dimensional Doppler Echocardiography
D066298 In Vitro Techniques Methods to study reactions or processes taking place in an artificial environment outside the living organism. In Vitro Test,In Vitro Testing,In Vitro Tests,In Vitro as Topic,In Vitro,In Vitro Technique,In Vitro Testings,Technique, In Vitro,Techniques, In Vitro,Test, In Vitro,Testing, In Vitro,Testings, In Vitro,Tests, In Vitro,Vitro Testing, In

Related Publications

Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
January 1993, Biomedical sciences instrumentation,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
August 2001, The Journal of thoracic and cardiovascular surgery,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
September 1989, The Annals of thoracic surgery,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
June 2020, The Journal of thoracic and cardiovascular surgery,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
January 1972, The Journal of thoracic and cardiovascular surgery,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
March 2021, The Journal of international medical research,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
November 1971, The British journal of surgery,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
January 2012, Computer methods in biomechanics and biomedical engineering,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
July 1995, The Journal of heart valve disease,
Morgane Evin, and Julien Magne, and Stuart M Grieve, and Régis Rieu, and Philipe Pibarot
July 2000, The International journal of artificial organs,
Copied contents to your clipboard!