In vitro pulsatile flow study on effective orifice area of prosthetic mechanical heart valves. 1993

G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
Baxter Healthcare, Inc., Edwards CVS Division, Irvine, CA 92714.

The effective orifice area (EOA) of a mechanical heart valve is an index of how well the valve design utilizes its primary orifice area (POA). In vitro measurements of EOA of aortic valves were maintained by means of pressure drop and root mean square flow rate measurements in a pulse duplicator during systole. Edwards-Duromedics, St. Jude Medical, and Carbomedics aortic valves of sizes 19 19 mm, 21 mm, 25 mm, and 27 mm were analyzed over a cardiac output range of 3 to 7 liters/min. The resultant ratios of EOA/POA were in the range of 0.6-0.8. A simplified equation suggested by the FDA was used in this study to calculate EOAs. To agree with original assumptions of the simplification, the entrance flow area (EFA) where upstream pressure is measured, must be large as compared to that of the test valves. If not, the formula can yield questionable results such as implying that the EOA can be larger than the POA (Walker P et al, 1992) [1]. This paper discusses the limitations in using such an equation. In conclusion, we suggest utilizing the parameter square root of 1-(POA/EFA)2 to evaluate the validity of the data processing, before using the equation. The parameter should be close to one, and in this study it was 0.997.

UI MeSH Term Description Entries
D011474 Prosthesis Design The plan and delineation of prostheses in general or a specific prosthesis. Design, Prosthesis,Designs, Prosthesis,Prosthesis Designs
D011673 Pulsatile Flow Rhythmic, intermittent propagation of a fluid through a BLOOD VESSEL or piping system, in contrast to constant, smooth propagation, which produces laminar flow. Flow, Pulsating,Perfusion, Pulsatile,Flow, Pulsatile,Flows, Pulsatile,Flows, Pulsating,Perfusions, Pulsatile,Pulsatile Flows,Pulsatile Perfusion,Pulsatile Perfusions,Pulsating Flow,Pulsating Flows
D006350 Heart Valve Prosthesis A device that substitutes for a heart valve. It may be composed of biological material (BIOPROSTHESIS) and/or synthetic material. Prosthesis, Heart Valve,Cardiac Valve Prosthesis,Cardiac Valve Prostheses,Heart Valve Prostheses,Prostheses, Cardiac Valve,Prostheses, Heart Valve,Prosthesis, Cardiac Valve,Valve Prostheses, Cardiac,Valve Prostheses, Heart,Valve Prosthesis, Cardiac,Valve Prosthesis, Heart
D001696 Biomechanical Phenomena The properties, processes, and behavior of biological systems under the action of mechanical forces. Biomechanics,Kinematics,Biomechanic Phenomena,Mechanobiological Phenomena,Biomechanic,Biomechanic Phenomenas,Phenomena, Biomechanic,Phenomena, Biomechanical,Phenomena, Mechanobiological,Phenomenas, Biomechanic

Related Publications

G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
November 2017, The Journal of heart valve disease,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
February 1966, The Journal of thoracic and cardiovascular surgery,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
June 2020, The Journal of thoracic and cardiovascular surgery,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
January 1983, Life support systems : the journal of the European Society for Artificial Organs,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
January 2017, The Journal of heart valve disease,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
January 1995, Journal of medical engineering & technology,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
December 1974, Thoraxchirurgie, vaskulare Chirurgie,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
August 2001, The Journal of thoracic and cardiovascular surgery,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
September 1990, Journal of biomedical engineering,
G X Guo, and C M Romero, and C Kingsbury, and R C Quijano
January 1979, Journal of biomechanics,
Copied contents to your clipboard!