Molecularly Imprinted Polymers Based Electrochemical Sensor for 2,4-Dichlorophenol Determination. 2016

Benzhi Liu, and Hui Cang, and Jianxiang Jin
School of Environmental Science and Engineering, Yancheng Institute of Technology, 224051 Yancheng, China. lbzycit@163.com.

A molecularly imprinted polymers based electrochemical sensor was fabricated by electropolymerizing pyrrole on a Fe₃O₄ nanoparticle modified glassy carbon electrode. The sensor showed highly catalytic ability for the oxidation of 2,4-dichlorophenol (2,4-DCP). Square wave voltammetry was used for the determination of 2,4-DCP. The oxidation peak currents were proportional to the concentrations of 2,4-DCP in the range of 0.04 to 2.0 µM, with a detection limit of 0.01 µM. The proposed sensor was successfully applied for the determination of 2,4-DCP in water samples giving satisfactory recoveries.

UI MeSH Term Description Entries

Related Publications

Benzhi Liu, and Hui Cang, and Jianxiang Jin
November 2013, The Analyst,
Benzhi Liu, and Hui Cang, and Jianxiang Jin
February 2020, Sensors (Basel, Switzerland),
Benzhi Liu, and Hui Cang, and Jianxiang Jin
May 2023, Mikrochimica acta,
Benzhi Liu, and Hui Cang, and Jianxiang Jin
January 2014, Analytica chimica acta,