Feedback control dynamics for glucose controlled insulin infusion system. 1979

A H Clemens

Miles Laboratories has developed a Glucose Controlled Insulin Infusion System (GCIIS) designated by the Trademark (BIOSTATOR) as a tool to investigate the physiologic control parameters of carbohydrate metabolism and regulatory deficiencies in diabetes. It consists, in principle, of a rapid on-line glucose analyzer, a computer/controller for the calculation and control of insulin or dextrose infusion, and a multichannel infusion system. A silent printer records, on a minute-by-minute basis, the glucose value measured, the insulin and/or dextrose infusion rates, and the cumulative total of the insulin infused. The on-line glucose analyzer employs an electrochemical sensor with immobilized glucose oxidase and measures the hydrogen peroxide produced. Its linearity extends beyond 700 mg/dl glucose; it permits a rapid two-point calibration of the sensor and the overall calibration of the on-line analyzer without disconnecting the catheter from the patient. The system's response time, including blood sample transport from the patient, is less than 90 seconds with a blood loss of approximately 50 ml per 24 h. The insulin and dextrose infusions are governed by control algorithms; various mathematical models have been developed and employed toward improved feedback control dynamics. The rapid glucose analyzer eliminates the need for the calculation of a "predicted" glucose value and permits, instead, the use of a derivative function for dynamic control. A multichannel infusion system combines the principles of a peristaltic pump with the advantages of a precision pump performance and individual computer control for each of the pump channels.

UI MeSH Term Description Entries
D007261 Infusions, Intra-Arterial Regional infusion of drugs via an arterial catheter. Often a pump is used to impel the drug through the catheter. Used in therapy of cancer, upper gastrointestinal hemorrhage, infection, and peripheral vascular disease. Infusions, Regional Arterial,Infusions, Intra Arterial,Infusions, Intraarterial,Arterial Infusion, Intra,Arterial Infusion, Regional,Arterial Infusions, Intra,Arterial Infusions, Regional,Infusion, Intra Arterial,Infusion, Intra-Arterial,Infusion, Intraarterial,Infusion, Regional Arterial,Intra Arterial Infusion,Intra Arterial Infusions,Intra-Arterial Infusion,Intra-Arterial Infusions,Intraarterial Infusion,Intraarterial Infusions,Regional Arterial Infusion,Regional Arterial Infusions
D007328 Insulin A 51-amino acid pancreatic hormone that plays a major role in the regulation of glucose metabolism, directly by suppressing endogenous glucose production (GLYCOGENOLYSIS; GLUCONEOGENESIS) and indirectly by suppressing GLUCAGON secretion and LIPOLYSIS. Native insulin is a globular protein comprised of a zinc-coordinated hexamer. Each insulin monomer containing two chains, A (21 residues) and B (30 residues), linked by two disulfide bonds. Insulin is used as a drug to control insulin-dependent diabetes mellitus (DIABETES MELLITUS, TYPE 1). Iletin,Insulin A Chain,Insulin B Chain,Insulin, Regular,Novolin,Sodium Insulin,Soluble Insulin,Chain, Insulin B,Insulin, Sodium,Insulin, Soluble,Regular Insulin
D008991 Monitoring, Physiologic The continuous measurement of physiological processes, blood pressure, heart rate, renal output, reflexes, respiration, etc., in a patient or experimental animal; includes pharmacologic monitoring, the measurement of administered drugs or their metabolites in the blood, tissues, or urine. Patient Monitoring,Monitoring, Physiological,Physiologic Monitoring,Monitoring, Patient,Physiological Monitoring
D001786 Blood Glucose Glucose in blood. Blood Sugar,Glucose, Blood,Sugar, Blood
D003201 Computers Programmable electronic devices designed to accept data, perform prescribed mathematical and logical operations at high speed, and display the results of these operations. Calculators, Programmable,Computer Hardware,Computers, Digital,Hardware, Computer,Calculator, Programmable,Computer,Computer, Digital,Digital Computer,Digital Computers,Programmable Calculator,Programmable Calculators
D003920 Diabetes Mellitus A heterogeneous group of disorders characterized by HYPERGLYCEMIA and GLUCOSE INTOLERANCE.
D005246 Feedback A mechanism of communication within a system in that the input signal generates an output response which returns to influence the continued activity or productivity of that system. Feedbacks
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man

Related Publications

A H Clemens
January 1979, Hormone and metabolic research. Supplement series,
A H Clemens
April 1984, Biomedizinische Technik. Biomedical engineering,
Copied contents to your clipboard!