Venturi pressure cannot cause cavitation in mechanical heart valve prostheses. 1991

J M Gross, and G X Guo, and N H Hwang
Department of Biomedical Engineering, Memphis State University, Tennessee 38152.

Surface pitting of certain mechanical heart valve (MHV) explants has prompted investigation into possible causes of cavitation during MHV operation. Leaflets of a 29 mm MHV were glued shut with B-datum (BD) gaps fixed at 0.0089, 0.0174, and 0.0219 cm. Each BD gap setting was tested in a steady flow chamber, with leakage flow established at transvalvular pressures (delta P) of 20 to 200 mmHg. Laser Doppler velocimeter (LDV) velocity measurements were recorded 220 microns distal to the BD, along with leakage flowrates. Maximum LDV velocities were compared with those calculated using the mass conservation equation. At identical P, the LDV flow velocities for the three BD settings were found to be approximately equal. This indicates a geometric independence of the leakage flow velocity. At atmospheric pressure, the local velocity necessary to cavitate blood as a liquid is approximately 13 m/sec. These results demonstrate that the leakage velocity is insufficient to cause cavitation. A simplified theoretical model is proposed to illustrate the necessary delta P to produce Venturi related cavitation.

UI MeSH Term Description Entries
D008955 Models, Cardiovascular Theoretical representations that simulate the behavior or activity of the cardiovascular system, processes, or phenomena; includes the use of mathematical equations, computers and other electronic equipment. Cardiovascular Model,Cardiovascular Models,Model, Cardiovascular
D011474 Prosthesis Design The plan and delineation of prostheses in general or a specific prosthesis. Design, Prosthesis,Designs, Prosthesis,Prosthesis Designs
D011475 Prosthesis Failure Malfunction of implantation shunts, valves, etc., and prosthesis loosening, migration, and breaking. Prosthesis Loosening,Prosthesis Durability,Prosthesis Migration,Prosthesis Survival,Durabilities, Prosthesis,Durability, Prosthesis,Failure, Prosthesis,Failures, Prosthesis,Loosening, Prosthesis,Loosenings, Prosthesis,Migration, Prosthesis,Migrations, Prosthesis,Prosthesis Durabilities,Prosthesis Failures,Prosthesis Loosenings,Prosthesis Migrations,Prosthesis Survivals,Survival, Prosthesis,Survivals, Prosthesis
D001783 Blood Flow Velocity A value equal to the total volume flow divided by the cross-sectional area of the vascular bed. Blood Flow Velocities,Flow Velocities, Blood,Flow Velocity, Blood,Velocities, Blood Flow,Velocity, Blood Flow
D006350 Heart Valve Prosthesis A device that substitutes for a heart valve. It may be composed of biological material (BIOPROSTHESIS) and/or synthetic material. Prosthesis, Heart Valve,Cardiac Valve Prosthesis,Cardiac Valve Prostheses,Heart Valve Prostheses,Prostheses, Cardiac Valve,Prostheses, Heart Valve,Prosthesis, Cardiac Valve,Valve Prostheses, Cardiac,Valve Prostheses, Heart,Valve Prosthesis, Cardiac,Valve Prosthesis, Heart
D006439 Hemodynamics The movement and the forces involved in the movement of the blood through the CARDIOVASCULAR SYSTEM. Hemodynamic
D006461 Hemolysis The destruction of ERYTHROCYTES by many different causal agents such as antibodies, bacteria, chemicals, temperature, and changes in tonicity. Haemolysis,Extravascular Hemolysis,Intravascular Hemolysis,Extravascular Hemolyses,Haemolyses,Hemolyses, Extravascular,Hemolyses, Intravascular,Hemolysis, Extravascular,Hemolysis, Intravascular,Intravascular Hemolyses
D006801 Humans Members of the species Homo sapiens. Homo sapiens,Man (Taxonomy),Human,Man, Modern,Modern Man
D012815 Signal Processing, Computer-Assisted Computer-assisted processing of electric, ultrasonic, or electronic signals to interpret function and activity. Digital Signal Processing,Signal Interpretation, Computer-Assisted,Signal Processing, Digital,Computer-Assisted Signal Interpretation,Computer-Assisted Signal Interpretations,Computer-Assisted Signal Processing,Interpretation, Computer-Assisted Signal,Interpretations, Computer-Assisted Signal,Signal Interpretation, Computer Assisted,Signal Interpretations, Computer-Assisted,Signal Processing, Computer Assisted

Related Publications

J M Gross, and G X Guo, and N H Hwang
March 1991, The International journal of artificial organs,
J M Gross, and G X Guo, and N H Hwang
October 1994, Artificial organs,
J M Gross, and G X Guo, and N H Hwang
January 2003, APMIS. Supplementum,
J M Gross, and G X Guo, and N H Hwang
January 1996, ASAIO journal (American Society for Artificial Internal Organs : 1992),
J M Gross, and G X Guo, and N H Hwang
September 2004, Expert review of medical devices,
J M Gross, and G X Guo, and N H Hwang
October 2004, The International journal of artificial organs,
J M Gross, and G X Guo, and N H Hwang
October 1991, Nederlands tijdschrift voor geneeskunde,
J M Gross, and G X Guo, and N H Hwang
January 1996, The Journal of heart valve disease,
J M Gross, and G X Guo, and N H Hwang
July 2004, Kyobu geka. The Japanese journal of thoracic surgery,
J M Gross, and G X Guo, and N H Hwang
March 2003, The Journal of heart valve disease,
Copied contents to your clipboard!